Corning Siecor OTDR Plus Multitester II
Model 340M-4210
with Hard Carrying Case
Model: 340M-4210
S/N: 2008-1267
Includes: Power Adapter SW101: Input: 100-250 V / 50-60 Hz / 1.0 - 0.5A Output: 12 V / 3.5 A.
Software Upgrade: V. 3.23C
Hard Carrying Case
Note: Will Include new replacement battery: Panasonic LCS 2312
This tester is in excellent operational condition, and will include a new replacement battery.
N&R Scientific Company, Inc. - also known as NR Scientific, is the owner and seller of the item offered for sale. Our primary warehouse and refurbishment facility is located on Roosevelt Place
in Palisades Park, New Jersey. We are
minutes from Manhattan, and local to all 3 New York City Area Airports. Items located at our primary facility in Palisades Park can generally be made available for acceptance testing/inspection prior to shipment.
We have been providing the highest quality refurbished and fully guaranteed equipment to customers worldwide for many years.
You can purchase with complete confidence that refurbished equipment will be provided in excellent cosmetic and perfect operational condition and that non-refurbished equipment is accurately described and represented.
Manufacturer's Description:
The Model 340 OTDR Plus
Multitester II offers true multi-testing capability that features field-installable single-mode, multimode and quad OTDR modules,visual fault locator, as well as optional built-in power meter and laser source. The standard unit provides internal solid-state memory for 125 traces and a floppy disk drive.An optional hard drive is also offered to allow mass data storage and speeds testing by eliminating the need to constantly exchange floppy disks in the standard drive.The product line offers a large variety of high-performance multimode and single-mode modules featuring a superior single-mode unit with 46 dB dynamic range. SoftView
Emulation Software provides desktop analysis and quality documentation capability.
Specifications:
Optical Modules
Multimode Modules 340M-40 340M-41 340M-42
Center Wavelength (±20 nm) 850 nm 1300 nm 850/1300 nm
Dynamic Range 23 dB 26 dB 23/26 dB
Attenuation Deadzone 6.5 m 7 m 6.5/7 m
Single-mode Modules 340M-13 340M-14 340M-15 340M-23 340M-24 340M-25
Center Wavelength 1310 nm 1550 nm 1310/1550 1310 nm 1550 nm 1310/1550
± 20 nm ± 20 nm 20/± 20 nm ± 20 nm ± 20 nm ± 20 nm
Dynamic Range 30 dB 28 dB 30/28 dB 36 dB 34 dB 36/34 dB
Attenuation Deadzone 10 m 12 m 10/12 m 10 m 12 m 10/12 m
Single-mode Modules 340M-34 340M-36 340M-38 340M-39 340M-61 340M-62
Center Wavelength (±20 nm) 1550 nm 1310/1550 nm 1550 nm 1310/1550 nm 1244 nm 1244/1310 nm
Dynamic Range 40 dB 40/40 dB 46 dB 43/46 dB 36 dB 36/36 dB
Attenuation Deadzone 6 m 6 m 10 m 10/10 m 6 m 6/9.5 m
Single-mode Modules 340M-63 340M-64 340M-71 340M-72 340M-73 340M-74 340M-76
Center Wavelength (±20 nm) 1244/1550 nm 1244/1625 nm 1625 nm 1310/1625 nm 1550/1625 nm 1625 nm 1550/1625 nm
Dynamic Range 36/34 dB 36/36 dB 36 dB 36/36 dB 36/36 dB 40 dB 40/40 db
Attenuation Deadzone 6/11 m 6/7 m 7 m 9.5/7 m 11/7 m 7 m 11/7 m
Hybrid Multimode/
Single-mode Modules 340M-56 340M-57 340M-53 340M-54
Center Wavelength (±20 nm) 850/1300 nm 850/300 nm 1310/1550/1625 nm 1310/1410 nm
1310/1550 nm 1310/1550 nm 1550/1625 nm
Dynamic Range 23/25 dB 23/24 dB 40/40/40 dB 36/36 dB
22/21 dB 33/31 dB 36/36 dB
Attenuation Deadzone 8/9 m 8/9 m 7/7/7 m 7/7 m
11/12 m 11/12 m 7/7 m
General Module
Parameter Specification
Universal 2.5 mm Connector Types3 FC, ST® Compatible, SC, D4, Biconic,
(Ultra PC standard) DIN 47256, SMA 905/906
(Angle PC available) Diamond HMS-0/HMS-10/HMS-10A, E-2000
Reflective Deadzone (typical) 3 m (multimode) / 3.5 m (single-mode)
Pulse width 4 ns to 10 µs (multimode); 10 ns to 30 µs (single-mode)
(wavelength dependent)
Spectral Width (RMS) 10 nm
Loss Resolution 0.001 dB
Distance Resolution 0.0001 km; 0.01 meters; 0.001 kft; 1ft; 0.0001 mi
Distance Sampling 0.25, 0.5, 1, 2, 4, 8, 16 m (range dependent)
Distance Accuracy 0.0025% of distance measurement +/- distance resolution +/- index uncertainty
Laser Certification CDRH class 1 requirements (eye safe) 21 CFR
(F-cartM6behindswitch)